Notre groupe organise plus de 3 000 séries de conférences Événements chaque année aux États-Unis, en Europe et en Europe. Asie avec le soutien de 1 000 autres Sociétés scientifiques et publie plus de 700 Open Access Revues qui contiennent plus de 50 000 personnalités éminentes, des scientifiques réputés en tant que membres du comité de rédaction.
Les revues en libre accès gagnent plus de lecteurs et de citations
700 revues et 15 000 000 de lecteurs Chaque revue attire plus de 25 000 lecteurs
Amadeo Pesce, Cameron West, Robert West, Sergey Latyshev, David Masters-Moore, Patrick Friel, John M. Hughes, Perla Almazan, Elizabeth Gonzales and Charles Mikel
Background: Laboratory urine drug testing of patients on chronic opioid therapy requires providing a large test menu of medications commonly prescribed for this population as well as metabolites and illicit substances. It has been shown that liquid chromatography-tandem mass spectrometry (LC-MS/MS) is the preferred method to analyze urine specimens for these substances.
Purpose of the study: To describe the challenges and some of the techniques to validate the analytical procedures used to identify and quantify these medications and substances.
Methods: Using data obtained from testing over one million specimens, the authors developed a proposed test menu. Potential isobaric interferences were established by using literature references. A list of potentially interfering medications was obtained by using the proposed test menu and the most commonly prescribed medications. Finally, criteria were designed to detect possible carryover.
Results: The LC-MS/MS instrumentation eliminated all potential interferences and provided quantitative data over the test range needed to monitor these patients. Carryover could be eliminated by setting the carryover thresholds for each analyte.
Conclusions: Reference laboratories utilizing LC-MS/MS technology to conduct urine drug testing for pain clinicians should employ specific techniques described in this study to develop an optimal test menu and validate procedures that include isolating retention times for isobaric compounds, identifying interfering substances including impurities in medicinal and illicit substance preparations, monitoring ion suppression, and avoiding carryover.